Hydrophilic Coating for Vascular Catheters

In short

Vascular catheter hydrophilic coatings must satisfy both low friction and haemocompatibility requirements, with haemolysis, platelet and coagulation testing support provided under ISO 10993-4.

At a Glance

Contact classNMPA Class III · Blood contact
Recommended seriesChangtu™ CV Series, Meitu™ MT Series
Typical substratesTPU, PA (nylon), PVC, Pebax
Key tests to plan
  • ISO 10993-4 / GB/T 16886.4 blood interaction evaluation
  • Haemolysis, platelet adhesion and clotting time assays
  • Coefficient of friction testing under simulated vascular conditions
  • Coating integrity verification under deformation

Contact class shown is that of the recommended coating series, not a classification of the device. The device's own class is determined by its registration certificate.

What this device demands

Haemocompatibility
Haemolysis, platelet adhesion and clotting time must be evaluated per ISO 10993-4
Low friction
Intravascular tracking requires a low coefficient of friction to reduce the risk of vessel wall damage
Compliance
The coating must withstand catheter deformation within the vessel without cracking
Protein adsorption
In blood contact, adsorption of plasma proteins at the coating surface must be assessed as part of the biological evaluation

Testing and standards to plan for

  • ·ISO 10993-4 / GB/T 16886.4 blood interaction evaluation
  • ·Haemolysis, platelet adhesion and clotting time assays
  • ·Coefficient of friction testing under simulated vascular conditions
  • ·Coating integrity verification under deformation

The final evaluation battery is determined by the device registration applicant based on contact nature and duration (ISO 10993-1). The list above is a planning input, not a substitute for that determination.

Common Questions

What are the key biological evaluation points for vascular catheter coatings?
For blood-contacting devices the focus is blood interaction, assessed per ISO 10993-4 (GB/T 16886.4 in China), typically covering haemolysis, thrombosis, platelet and coagulation endpoints. The test battery is determined by the device registration applicant based on contact nature and duration — short-term under 24 hours, for example, or prolonged. As a material component, the coating must be supported by corresponding material characterisation and test data.
Does protein adsorption need to be considered for vascular catheter coatings?
Vascular catheters are in blood contact, so plasma protein adsorption at the coating surface is part of the overall biological evaluation and must be considered alongside haemolysis, platelet and coagulation endpoints. Formulation design involves a trade-off between protein adsorption behaviour and low friction; the priority depends on the device's functional positioning and should be settled with comparative data measured under actual use conditions.
Will the coating crack when the catheter bends?
Coating flexibility depends on crosslink density, film thickness and substrate compatibility. Vascular catheters deform during tracking, so coating integrity should be verified under deformation at the selection stage; we can support with a fit recommendation.

Recommended series

Discuss your device

Send the device construction, substrate grade and target performance. We will confirm whether the recommended series fits and arrange samples where appropriate.