How is hydrophilic coating friction measured?
In China, the governing method is YY/T 1898-2024, Test method for the durability of hydrophilic coatings on intravascular catheters and guidewires, issued 7 February 2024 and effective 1 March 2025. It defines four test methods — appearance, friction force, particulate, and chemical performance — of which the friction force method is the core quantitative approach: after hydration in physiological saline, the sample is held between two silicone pads at constant normal force and reciprocated at 200–600 mm/min at (37±2) °C for at least 10 cycles, with friction force and coefficient of friction recorded as a curve, then the mean of the first five cycles compared with the last five.