Coating Solutions by Device Type

In short

Hydrophilic coating selection starts from the device, not the coating. The two decisive factors are whether the device contacts blood and what the substrate is; the third is whether the production line can support the required cure process. The pages below set out those requirements for the four device categories Medcoa works with most often.

Hydrophilic Coating for Urinary Catheters

Coating solution for indwelling and intermittent urinary catheters, combining lubricity with biological evaluation documentation.

Requirements and tests →

Hydrophilic Coating for Vascular Catheters

Coating solution for blood-contacting vascular catheters, covering haemocompatibility documentation.

Requirements and tests →

Coating for Balloon Catheters

Coating solution for thin-walled substrates such as PA balloons, focused on coating uniformity and compliance.

Requirements and tests →

Coating for Guidewires

Coating solution for interventional and steerable guidewires, focused on uniform coverage at very small diameters and flexibility.

Requirements and tests →

Common Questions

Selection and testing questions by device category.

Which coating should I use for urinary catheters?
Urinary catheters are generally non-blood-contact Class II devices on TPU or PVC shafts, so Youtu™ UC is usually appropriate. If the product has specific particulate release or lubricity retention requirements, provide the device construction and target performance and the series and process can be recommended accordingly.
What testing is required for blood-contacting catheters?
Blood-contacting devices are evaluated primarily for blood interaction under ISO 10993-4 (GB/T 16886.4 in China), typically covering haemolysis, thrombosis, platelet and coagulation endpoints, alongside the basic ISO 10993 series endpoints. The final battery is determined by the device registration applicant based on contact nature and duration. Particulate release from the coating is assessed separately.
Can balloon catheters be coated without affecting dimensions?
Thin-walled balloons demand tight control of coating uniformity and film thickness, and the effect on nominal balloon size and trackability must be limited. PA has relatively low surface energy and generally needs stronger pre-treatment than TPU, such as plasma activation or a primer. Coating integrity should be verified after inflation cycling on actual balloon samples.
What is different about guidewire coating?
Guidewires combine a very small diameter with a high aspect ratio, so leveling and drainage during dip coating is hard to control and local thickness variation readily appears. Adjusting coating viscosity, withdrawal speed and cure process — or switching to spray coating — is usually needed. Guidewires are blood-contacting, so ISO 10993-4 endpoints apply, and the transition between hydrophilic and hydrophobic segments must be smooth.

Device not listed?

Sheaths, introducers and other interventional or drainage devices are coated on the same platform. Describe the device construction, substrate and contact class, and we will assess fit and recommend a series.