How is hydrophilic coating friction measured?
In short
In China, the governing method is YY/T 1898-2024, Test method for the durability of hydrophilic coatings on intravascular catheters and guidewires, issued 7 February 2024 and effective 1 March 2025. It defines four test methods — appearance, friction force, particulate, and chemical performance — of which the friction force method is the core quantitative approach: after hydration in physiological saline, the sample is held between two silicone pads at constant normal force and reciprocated at 200–600 mm/min at (37±2) °C for at least 10 cycles, with friction force and coefficient of friction recorded as a curve, then the mean of the first five cycles compared with the last five.
There is now a standard to follow
YY/T 1898-2024 was issued by the National Medical Products Administration on 7 February 2024 and took effect on 1 March 2025. It is China's first industry standard dedicated to test methods for the durability of hydrophilic lubricious coatings on intravascular catheters and guidewires.
The standard applies to intravascular catheters and guidewires carrying a hydrophilic lubricious coating. Its scope is explicitly bounded: it covers only polymer-based hydrophilic lubricious coatings chemically bound to the substrate surface, and excludes drug-loaded coatings, biodegradable coatings and other types. It cannot be applied directly to the evaluation of antibacterial or biodegradable coatings.
The standard references GB/T 14233.1, GB/T 16886.12, GB/T 6682, ISO 10993-4, YY/T 0807 and YY/T 1556, and links particulate counting to method 0903 (insoluble particulate matter) of the Chinese Pharmacopoeia, 2020 edition, Volume IV.
Definition
Coating fastness (fastness of the hydrophilic coating)
The ability of a coating to resist loss and retain lubricity under service conditions such as friction and immersion. It is not a single metric but a system reflected at different scales through four test types: appearance, friction force, particulates and chemical performance.
How the four methods divide the work
The standard's core logic is scale-separated evaluation: appearance catches macroscopic defects, friction force tracks lubricity decay, particulates count what has been shed, and chemical performance captures dissolution or nanoscale migration. The four methods complement each other across levels from the visible down to sub-micron.
| Method | Scale | What is observed | Where it is used |
|---|---|---|---|
| Appearance (staining, Annex A) | Macroscopic / optical | Immersion for about 1 min in 1% Congo red or 0.1% basic fuchsin, which stains the coating but not the substrate; examination under an optical microscope at 10× or greater for coating loss or cracking | First-pass screening of coating integrity and uniformity |
| Friction force (Annex B) | Mechanical / dynamic | Constant normal force applied through silicone pads while the sample is lifted vertically; friction force and coefficient of friction recorded across cycles and plotted as a curve | Lubricity decay and durability — the core quantitative method |
| Particulate (Annexes D/E) | Particle / count | Test liquid prepared by immersion (Annex D) or vascular model friction (Annex E) according to clinical use, then particle size and count determined per Pharmacopoeia 0903 or YY/T 1556 | Particulate risk from coating loss |
| Chemical performance (Annex F) | Molecular / chemical | Loss components that dissolve or form sub-micron and nanoscale particles, characterised by reducing substances, evaporation residue, UV absorbance or other specific methods | The fraction of loss that particle counting cannot detect |
Parameters of the friction force test
This section covers the specific condition ranges that can be cited from the standard. In practice they must be confirmed against the sample construction, then fixed and written into the method file. The standard also requires a friction force / coefficient of friction curve to monitor change in real time, rather than recording a single number.
The standard defines one friction cycle precisely: starting from the position where the silicone pads clamp the sample under constant normal force, sliding a specified distance and releasing back to the start. The number of cycles should be chosen with statistical significance in mind; the standard recommends at least 10, and this may be adjusted to the device's intended clinical use.
- ·Avoid twisting or re-clamping during the run, which introduces mechanical damage and corrupts force readings
- ·Insert a support mandrel into thin-walled catheters to prevent wall deformation under clamping and friction
- ·Keep the two silicone pads parallel, clean and scratch-free; the water level must cover the pads
- ·After the run, remove the sample, inspect coating integrity in the clamped region and record it
| Item | Condition or requirement | Note |
|---|---|---|
| Medium and temperature | Physiological saline, (37±2) °C water bath | Simulates physiological conditions and activates the hydrophilic coating |
| Normal clamping force | 2–5 N, low enough that the sample shows no obvious deformation | More force is not better; confirm against catheter or guidewire wall thickness and construction |
| Reciprocating speed | 200–600 mm/min | Corresponds to the order of magnitude of clinical insertion rates |
| Number of cycles | At least 10 recommended | Consider statistical significance; adjust to clinical use |
| Test distance | Generally not less than 30 mm | Set according to coated length |
| Immersion time | Generally not less than 60 s | Ensure the test section is fully wetted; clamp quickly after removal to avoid drying |
| Counterface | Two silicone pads | Clamping force control accuracy and pad specification must meet the standard |
| Sample preparation | Cut and discard samples with scratches or damage | A support mandrel may be inserted into thin-walled catheters to prevent deformation |
How the friction ratio is calculated
The standard provides a trend indicator rather than a single pass/fail line: take the mean friction force (or coefficient of friction) of the last five cycles and of the first five, and compute the ratio to quantify how much lubricity changed over repeated friction.
When reading the curve, judge the shape of the trend rather than any absolute value. An acceptable curve is broadly flat, fluctuates slightly as cycles accumulate, shows no sudden upward step and has no severe signal spikes. A steep rise in some region, a large jump in force or a marked increase in spikes usually indicates coating wear and loss, substrate exposure and loss of lubricity.
The standard does not set an acceptance limit; that is defined by the product's technical requirements. Common practice is to set an upper limit on the relative rise in coefficient of friction or friction force after a number of cycles, combined with visual assessment for blistering, flaking or shedding after the test.
Definition
Coefficient of friction (μ)
The ratio of friction force to normal clamping force, μ = F/N. Both static and kinetic coefficients exist. The value varies with temperature, medium, speed, clamping force and counterface material, so every report must state the full test conditions — a value without them is meaningless.
Reporting and verification
The standard's test report clause requires the test conditions and results to be listed, including sample number and results. Once parameters are fixed, replicate testing is needed for repeatability, with the number of replicates specified in the method file in advance — not written in after testing.
Equipment must meet the standard's requirements for clamping force accuracy and silicone pad specification, and the laboratory should complete IQ/OQ/PQ qualification, with attention to friction sensor calibration, cleanliness control of the particulate testing environment and selectivity verification of chemical methods.
Three interface relationships are often overlooked. When using the vascular model friction method to prepare test liquid, the model should meet ASTM F2394 or YY/T 0807. Immersion test conditions are given in YY/T 1556. Test water must be GB/T 6682 grade 2 or better, or other water filtered through a membrane no larger than 0.22 μm.
Method selection should follow clinical use: neurovascular catheters favour a combination of friction force and particulate testing, coronary guidewires favour appearance plus chemical performance, and peripheral vascular stent delivery systems are usually tested with all four methods.
How to use a supplier's data
A supplier can provide data measured by this method, but the boundary of applicability matters: the standard addresses intravascular catheters and guidewires. Non-vascular catheters such as urinary catheters differ in device type, counterface and clinical scenario, so the method must be adapted.
Results obtained on a supplier's standard substrate do not directly represent performance on a customer's actual device, because substrate grade, device construction, wall thickness and cure process all affect the outcome. Selection should be based on re-testing the customer's real samples, with test conditions and acceptance criteria written into the technical agreement.
Common Questions
- Which standard governs hydrophilic coating friction testing?
- For intravascular catheters and guidewires, the current basis is YY/T 1898-2024, Test method for the durability of hydrophilic coatings on intravascular catheters and guidewires, issued 7 February 2024 and effective 1 March 2025. The friction force method within it is the core quantitative approach. Note that the standard covers only polymer-based hydrophilic lubricious coatings chemically bound to the substrate; drug-loaded and biodegradable coatings are outside its scope.
- What parameters does the friction force test require?
- Physiological saline at (37±2) °C; normal clamping force 2–5 N, low enough to avoid obvious deformation; reciprocating speed 200–600 mm/min; at least 10 cycles recommended; test distance generally not less than 30 mm; immersion time generally not less than 60 s; two silicone pads as counterface. Parameters must be confirmed against sample construction, then fixed and kept consistent across reports and replicate tests.
- What coefficient of friction counts as a pass?
- YY/T 1898-2024 does not set an absolute pass value. It defines a trend evaluation: the ratio of mean friction force (or coefficient of friction) over the last five cycles to that over the first five, capturing the change after repeated cycles. Acceptance criteria come from the product's technical requirements — commonly an upper limit on relative rise, combined with assessment for blistering, flaking or shedding after the test.
- What else needs testing besides friction?
- The standard defines four methods: appearance (staining, optical microscopy for coating integrity), friction force (dynamic lubricity decay), particulate (counting shed particles per Pharmacopoeia 0903 or YY/T 1556) and chemical performance (characterising dissolved or nanoscale loss). Chemical testing covers the fraction that particle counting cannot detect. The four operate at different scales and should be used in combination.
- Can YY/T 1898-2024 be applied to urinary catheters?
- Not directly. Its scope is intravascular catheters and guidewires, and the evaluation logic is based on a simulated intravascular wet environment with silicone pad counterface. Urinary catheters contact the urethral environment, with different counterface, construction and intended use, so the method must be adapted. Supplier data obtained on urinary catheters should be verified separately against the applicable standard for non-vascular catheters.
- Can results from different laboratories be compared directly?
- Even under the same standard method, results are affected by clamping force, speed, pad batch, temperature control accuracy of the medium and sample handling. To compare coatings from different suppliers, unify method and parameters, run replicate tests in one laboratory, and compare curve shape rather than a single value alone.