Hydrophilic Coating for Urinary Catheters
In short
A urinary catheter hydrophilic coating forms a stable hydrated layer on a TPU or PVC shaft to reduce insertion friction, and is supplied with ISO 10993 series biological evaluation documentation to support registration.
At a Glance
| Contact class | NMPA Class II · Non-blood contact |
|---|---|
| Recommended series | Youtu™ UC Series |
| Typical substrates | TPU, PA (nylon), PVC |
| Key tests to plan |
|
Contact class shown is that of the recommended coating series, not a classification of the device. The device's own class is determined by its registration certificate.
What this device demands
- Lubricity retention
- Indwelling use requires the hydrated layer to maintain lubricity for a sufficient period in physiological media
- Adhesion
- The coating must not delaminate or peel after repeated bending and advancement of the catheter
- Particulate control
- Coating loss can generate particulates and must be evaluated and controlled
- Biocompatibility
- Mucosal contact requires the corresponding biological evaluation endpoints
Testing and standards to plan for
- ·Coefficient of friction testing (must state load, speed, medium and test duration)
- ·Lubricity retention testing (must state ambient temperature, humidity and test duration)
- ·Coating adhesion and durability testing
- ·ISO 10993 series biological evaluation (cytotoxicity, irritation, sensitisation and related endpoints)
The final evaluation battery is determined by the device registration applicant based on contact nature and duration (ISO 10993-1). The list above is a planning input, not a substitute for that determination.
Common Questions
- What testing does a urinary catheter hydrophilic coating require?
- Usually three groups. Performance testing covers coefficient of friction, lubricity retention, adhesion and durability. Biological evaluation follows the ISO 10993 series (GB/T 16886 in China), covering cytotoxicity, irritation and sensitisation. Particulate shedding from the coating is assessed separately. The final battery depends on the nature and duration of patient contact and should be determined by the device registration applicant per ISO 10993-1.
- What regulatory support documents can a coating supplier provide?
- We provide material chemical composition and safety data, material characterisation data, biological evaluation data performed under the ISO 10993 series (GB/T 16886 in China), and a TÜV SÜD biological evaluation report for hydrophilic-coated urinary catheters (Report No. 748267139301C/CN). NMPA Master File registration is not yet completed; we will inform customers once it is. If your submission timeline is fixed, confirm the document list and delivery schedule with us at project kick-off.
- Do TPU and PVC shafts need different coating solutions?
- Yes. The two substrates differ in surface energy, polarity and flexibility, so pre-treatment and formulation need to be adjusted accordingly. We recommend sending actual shaft samples for a fit test, with adhesion and coefficient of friction measured on the real substrate.
Recommended series
Discuss your device
Send the device construction, substrate grade and target performance. We will confirm whether the recommended series fits and arrange samples where appropriate.