Medcoa Biotechnology (Shanghai) Co., Ltd. logo
Base function

Hydrophilic Lubricity

In short

Hydrophilic lubricity is the base function of a hydrophilic coating. The coating polymer takes up water and swells, forming a hydrated layer on the device surface that converts dry friction against tissue into low-friction sliding in an aqueous medium.

How it works

Hydrophilic coatings use a hydrophilic polymer — polyvinylpyrrolidone (PVP) is the common choice — as the functional component. After coating and curing, the polymer forms a crosslinked network on the device surface. Once the device enters an aqueous or physiological environment, the network takes up water and swells into a hydrated layer, turning solid-solid contact into solid-liquid-solid lubrication. Lubricity retention depends on crosslink density and the stability of the hydrated layer, and these two parameters must be balanced during formulation: too low a crosslink density causes excessive water uptake and poor mechanical strength; too high restricts swelling and weakens lubricity.

Key points

Functional component
Hydrophilic polymer (PVP and similar)
Activation
Requires wetting in an aqueous or physiological environment
Key trade-off
Crosslink density vs. water uptake capacity
Typical supply form
Emulsion; our products are liquid, ready to use

How it is evaluated

Lubricity is quantified by the coefficient of friction (CoF). The current industry standard for coating durability on intravascular catheters and guidewires is YY/T 1898-2024 (Test method for the durability of hydrophilic coatings on intravascular catheters and guidewires, effective 2025-03-01). Its friction test applies a constant normal clamping force through a silicone pad and reciprocates the sample in physiological saline at (37±2) °C while recording the friction force and CoF curve. CoF depends strongly on clamping force, reciprocating speed, counterface material and medium, so any reported value must state its test conditions — otherwise it cannot be reproduced or compared. Durability is assessed from the change in performance and coating integrity after repeated friction cycles.

Where it applies

  • Vascular catheters
  • Interventional guidewires
  • Urinary catheters
  • Balloon catheters

Common Questions

What determines how lubricious a hydrophilic coating is?
Three things, together: the formulation, including the type and content of the hydrophilic component; the crosslink density, which sets the balance between hydrated-layer strength and water uptake; and the coating and curing process, including substrate pre-treatment, number of layers and cure conditions. No single factor determines final performance. Selection should be based on sample data measured on your actual substrate and process.
Why must a coefficient of friction always be reported with test conditions?
CoF is not an intrinsic material constant — it is a process-dependent quantity that changes with test conditions. A higher clamping force changes contact stress, reciprocating speed affects how the hydrated layer forms and breaks down, counterface material determines the contact mode, and the medium directly sets the lubrication regime. Without these, a value can be neither reproduced nor compared across products. YY/T 1898-2024 gives reference conditions of 2–5 N clamping force, 200–600 mm/min, (37±2) °C physiological saline and at least 10 reciprocating cycles; the specific values must be agreed against the sample construction and written into the technical agreement.
How is adhesion between the coating and the substrate ensured?
Adhesion depends on substrate surface energy and pre-treatment. TPU, PA, PVC, silicone and Pebax differ considerably in surface energy, so pre-treatment (plasma treatment or a primer, for example) must be selected accordingly and verified by adhesion testing. Metal substrates usually need a dedicated primer system. The process window is set against your substrate grade and device construction — sending samples for a fit test is the reliable route.
Does lubricity decline during prolonged use?
The hydrated layer gradually wears under continuous friction and fluid shear. The rate depends on crosslink density, coating thickness and the actual use environment. For devices with long procedure times, lubricity retention time or the performance retention rate after friction cycling are usually the right selection metrics. The specific decay curve has to come from testing that simulates actual use conditions.